Qingdao Daowan Technology Co., Ltd.——Daowan DW17 Multiparameter Water Quality Probe
Water Quality Monitoring | Multiparameter
Daowan DW17 Multiparameter Water Quality Probe
Daowan DW17 Multiparameter Water Quality Probe | Model: DW17/DW17-D
Product Positioning
The Daowan DW17 multiparameter water quality sonde is a multiparameter integrated observation instrument centered on CTD (conductivity, temperature, depth) measurement: the main unit houses high-precision conductivity, temperature and pressure (depth) measuring cells, with two sensor expansion slots allowing up to two additional sensors to be integrated simultaneously, so water-quality or optical parameters can be extended as required.

Fig. 1 Daowan DW17 Multiparameter Water Quality Probe (product photo)
Product Overview
The Daowan DW17 multiparameter water quality sonde is a multiparameter integrated observation instrument centered on CTD (conductivity, temperature, depth) measurement: the main unit houses high-precision conductivity, temperature and pressure (depth) measuring cells, with two sensor expansion slots allowing up to two additional sensors to be integrated simultaneously, so water-quality or optical parameters can be extended as required. The sonde uses a high-strength titanium alloy housing rated to 6,000 m. The DW17 standard configuration is the direct-reading & self-recording combined version with built-in rechargeable lithium battery and storage, suited to profiling and various monitoring systems; a direct-reading model (DW17-D) is also available, in which power and data acquisition are user-provided, for professional users with operational or integration experience.
The DW17 supports customized sensor configurations. Mature configurations successfully integrated to date include CTD + turbidity, CTD + pH, CTD + CDOM (colored dissolved organic matter), CTD + dissolved oxygen, and CTD + chlorophyll + turbidity (dual additional sensors), with monitoring elements flexibly extendable per user needs. Standardized data output formats ensure compatibility with mainstream data acquisition platforms and analysis software — a one-stop solution for customized monitoring needs.
Key Features
● High accuracy, high resolution, high stability, low power consumption;
● Hatch-free design: communication, power (charging) and data readout all without opening the housing;
● Easy to operate: plug-and-play cables, simple software interface, optional handheld terminal — no learning curve;
● Supplied with dedicated software supporting real-time display, graphics, parameter settings, battery status and more;
● The factory is equipped with a full set of calibration equipment; every instrument is calibrated before shipment to ensure accurate, reliable data;
● Supports remote upgrades for easy feature expansion and system maintenance, reducing cost of ownership;
● Core technologies independently controlled, with rapid after-sales response, remote guidance and calibration services.
Handheld Terminal (Optional)
The Daowan DW9 handheld terminal is a portable acquisition terminal supporting real-time sensor data acquisition, storage and display, running independently without a computer. It automatically identifies the connected device and sensor types — plug and play, making field data reading and logging easy. IP67 rated against water and dust, with 8 GB of built-in storage and 1 Hz sampling, it logs data continuously for 14 hours — an efficient, reliable mobile data acquisition solution for field personnel.
![]() Fig. 2 Daowan DW9 Handheld Terminal (optional) |
Host Software (Daowan DWS)
The equipment is managed by the Daowan DWS host software: on connection it automatically identifies device and sensor types, displaying each channel's measured values and working status in real time; measurement data are presented as profile plots, time series, T-S diagrams and real-time tables, with open, import and export support; the software provides device configuration (baud rate, Modbus address, sampling time and mode, special sampling), calibration coefficient management, data backup export (CSV) and firmware upgrades (online/manual), with a Chinese interface for easy operation.
![]() Fig. 3 DWS software real-time monitoring interface |
Technical Specifications
Main Unit Specifications
| Model | DW17 | DW17-D |
| Operating mode | Direct-reading / direct-reading & self-recording combined | |
| Instrument depth rating | Housing depth rating: 1,500 m / 6,000 m; overall system rating: subject to sensor rated depths | |
| Main unit dimensions | Main body φ50 mm × 450 mm, excluding external parts | Main body φ54 mm × 450 mm, excluding external parts |
| Weight | Main unit 1.6 kg, plus the weight of each sensor | Main unit 2.0 kg, plus the weight of each sensor |
| Housing material | Titanium alloy | |
| Sampling rate | max. 10 Hz, adjustable | |
| Communication | RS232 or RS485 | |
| Power supply | Direct-reading units: external power supply 12–24 V DC Direct-reading & self-recording combined units: built-in rechargeable lithium battery | |
| Memory | 16GB | |
Sensor Specifications
| Sensor | Rated depth | Measurement method | Maximum measuring range | Measurement accuracy | Resolution |
| Temperature | 6000m | Thermistor | -2~36℃ | ±0.002℃ | 0.0001℃ |
| Pressure | 6000m | Piezoresistive | 0~6000dbar* | ±0.05%F.S. | 0.002%F.S. |
| Conductivity | 6000m | Electromagnetic induction | 0~70mS/cm | ±0.005mS/cm | 0.0001mS/cm |
| pH | 100m | Electrode method | 0~14pH | ±0.1pH | 0.05pH |
| 1200m | Electrode method | 0~14pH | ±0.05pH | 0.003pH | |
| 6000m | Electrode method | 0~14pH | ±0.05pH | 0.01pH | |
| Chlorophyll | 100m | Fluorescence method | 0~200μg/L | ±2%F.S. | 0.01μg/L |
| 6000m | Fluorescence method | 0~150μg/L | ±2%F.S. | 0.02μg/L | |
| Dissolved oxygen | 100m | Fluorescence method | 0–20 mg/L or 0–200% Sat. | ±0.1mg/L | 0.05 mg/L or 0.5% Sat. |
| 100m | Electrochemical method | 0~150%Sat. | ±2%Sat. | 0.1%Sat. | |
| 6000m | Amperometric method | 0~150%Sat. | ±2%Sat. | 0.1%Sat. | |
| Turbidity | 100m | Light scattering method | 0~1000FTU | ±2%F.S. | 0.1FTU |
| 6000m | Light scattering method | 0~1250FTU | ±2%F.S. | 0.1FTU | |
| Blue-green algae | 100m | Fluorescence method | 0~100μg/L | ±3%F.S. | 0.05μg/L |
| Other sensors: ORP, PAR, CDOM, H2, H2S, rhodamine, fluorescein, CO2, O3, etc. | |||||
*Notes on pressure sensor measuring range
(1) The pressure sensor measuring range must be determined at ordering. Once selected, it cannot be changed.
(2) Optional measuring ranges: 0–50, 0–100, 0–200, 0–500, 0–1,000, 0–2,000, 0–4,000, 0–6,000 dbar. 1 dbar ≈ the pressure at 1 m water depth.
(3) Select a measuring range above and close to the maximum operating depth. (Example: for a maximum depth of 800 m, choose the nearest range above 800 dbar, i.e. 0–1,000 dbar)
(4) Overrange use is strictly prohibited. Damage caused by overrange use is not covered by the warranty.
Applications
● The DW17 standard configuration is the direct-reading & self-recording combined version with built-in rechargeable lithium battery and storage, suited to profiling and various monitoring systems;
● A direct-reading model is also available, in which power and data acquisition are provided by the user; suited to professional users with operational or integration experience.
Product Gallery
![]() Fig. 4 Daowan DW9 Handheld Terminal (optional) | ![]() Fig. 5 Delivered unit photo (CTD + pH sensor configuration) |
![]() Fig. 6 Assembled unit photo (CTD + CDOM sensor configuration, rear view) | ![]() Fig. 7 Assembled unit photo (CTD + CDOM sensor configuration, side view) |
![]() Fig. 8 Assembled unit photo (CTD + chlorophyll/turbidity + DO sensor configuration, front-side view) | ![]() Fig. 9 Assembled unit photo (CTD + chlorophyll/turbidity + DO sensor configuration, top view) |
![]() Fig. 10 Assembled unit photo (CTD + turbidity sensor configuration, side view) | ![]() Fig. 11 Close-up of the turbidity sensor tip (CTD + turbidity sensor configuration) |
![]() Fig. 12 Assembled unit photo (CTD + pH sensor configuration, mounting bracket, oblique side view) | ![]() Fig. 13 Assembled unit photo (CTD + pH sensor configuration, mounting bracket, top view) |